GD-270A Lubricant Grease Dropping Point Tester - GD-270A Specificaton & Trade Terms
Model | GD-270A |
---|
Place Of Origin | Chonqing |
---|
Price Term | EX-Work |
---|
Payment Term | T/T |
---|
The instrument is designed and made as per the Industry Standard of Petrochemical SH/T0115 “Test Methods for Dropping Point of Lubricating Grease and Solid Hydrocarbon”. The instrument is an air bath, and it is used to determine dropping point of lubricating grease and hydrocarbon over 100 °C.
Method Summary
Fill lubricating grease or solid hydrocarbon into a grease cup with a dropping point thermometer. The temperature when the first drop of sample falls from the grease cup or sample of 25 mm flows from test cup under the stipulated heating condition. The temperature is dropping point of the sample.
Technical Parameter
1. Power supply: AC 220V±10%, 50 Hz;
2. Heating power of electric furnace: 250W×2;
3. Heating control: Controlled by bidirectional silicon controlled rectifier;
4. Ambient temperature: ≤35°C;
5. Relative humidity: ≤85%;
6. Total power consumption: not more than 700 W;