Penetrating Coating Ultrasonic Thickness Gauge - UM-1D Specificaton & Trade Terms
Model | UM-1D |
---|
Place Of Origin | China |
---|
Price Term | EX-Work |
---|
Payment Term | T/T |
---|
UM-1D Penetrating Coating Ultrasonic Thickness Gauge
Standard package:
UM-1D thickness gauge
Standard probe (PT-08)
Carrying case
Two 1.5V AA alkaline batteries
Couplant
Operating manual
Certification of quality
Packaging list
Operating principle:
Standard mode: P-E (pulse-echo) method with dual-element probes
Coating mode: E-E (echo-echo) method with dual-element probes
Probe zero adjustment:
Selectable:
1-point calibration: normal measurement
2-point calibration: curve surface measurement or other special application with a high precision
V-path correction:
Automatic
Measuring range:
0.8mm to 300mm; 0.031in to 11.81in in standard mode
3.0mm to 18.0mm; 0.118in to 0.709in in through coating mode
Depending on probe, material and surface condition
Digital resolution:
0.001in or 0.1mm
Units:
Selectable: inch or mm
Display:
128×64 dot-matrix LCD screen (1.65in×2.24in; 42mm×57mm)
EL backlight and switchable contract ratio
Digit height: 0.54in or 13.75mm
Measurement update rate:
4 Hz in standard measurement mode
25 Hz in MIN capture mode
Material velocity range:
1000 to 9999 m/s; 0.039 to 0.394 inch/μs
Power:
Two 1.5V AA alkaline batteries
Warning with low battery voltage
Operating time:
Up to 200 hours (without backlight) with alkaline batteries, depending on operating mode
Auto shut-off:
After 5 minutes of non-use
Operating temperature:
-10 degree Celsius to +50 degree Celsius; +14 degree Fahrenheit to +122 degree Fahrenheit
Specification to -20 degree Celsius/-4 degree Fahrenheit on request
Size:
5.86in×2.87in×1.25in; 149mm×73mm×32mm (H×W×D)
Weight:
200g; 7oz (including batteries)
Data logger capacity:
Up to 500 readings can be divided into a maximum of 5 files (user-selectable)
Optional accessories:
Attachable rubber case
Numerous special probes on request
Couplants and high-temperature couplant
4-step/6-step block for calibration
Data transmission cable
Application software
Operating language:
English and Chinese
Measuring error:
Up to 1in: ±0.002in or ±0.1mm
Above 1in: ± 1%H
H is the height of the testing object
Repeatability:
±0.001in or ±0.1mm
Application software:
UMVIEW
Interfaces:
Serial RS 232C port in combination with a communicative cable;
Data transfer parameters: 8 bits, 1 stop bit, 9600 baud, no parity.
Main Products: Portable Ultrasonic Flaw Detector, Ultrasonic Thickness Gauge, Penetrating Coating Ultrasonic Thickness Gauge, Portable Hardness Tester, Pen-Type Hardness Tester, Digital Shore Durometer, Shore Durometer.
Key Words: NDT Instrument, Testing Instrument, Testing Equipment, Test Equipment, Testing Tools.
Please contact with us for more related productions!