puls Electrolyte Analyzer - CMS3300 Specificaton & Trade Terms
Model | CMS3300 |
---|
Place Of Origin | China |
---|
Packaging | 10pcs per box |
---|
Brand | Contec |
---|
Gurantee | one year |
---|
Certifications | CE,ISO |
---|
Price Term | EX-Work |
---|
Payment Term | T/T,Paypal |
---|
Supply Ability | 30,000pcs per week |
---|
Minimum Order | 10 pcs |
---|
Loading Ports | Shenzhen city,China |
---|
Delivery Time | 3 days |
---|
Features
CMS3300 Plus is a completely automatic Electrolyte Analyzer,whose applicable electrolyte
includes K/Na/Cl/iCa/nCa/TCa/pH/Li/TCO2 and Arion Gap(AG).It implements advanced
ion-selective electrode measurement technology to directly measure the concentration
of K/Na/Cl/iCa/nCa/TCa/pH/Li/TCO2 and Arion Gap(AG)in blood serum,blood palsma ,
whole blood ,and cerebrosipnal fluid and dilute urine.It is a kind of clinical instection
equipment which is quick accurate ,convenient and practical.
n The whole running process of the equipment is controlled by a computer . Also,
automatic electric potential tracking and correcting software are adopted to ensure its
stable performance.
n 320*240 The machine is featured with ultra large touchable screen with high
resolution ,human-machine interactive menu,and falut automatic alarming
nandremoving All these features can greatly improve work efficiency.
n The machine is free in switching between English and Chinese reports, meeting
requirements of different customers.
n The machine adopts innovative international patent technology such as wave
theory flushing method and direct flushing pipe method to avoid block and crossed
contamination.
n Every time putting the specimen into the machine ,the machine could simultaneously
detect eight ions including K/Na/Cl/iCa/nCa/TCa/pH/ Li/TCO2 and AG as well as
parameters.
nThe equipment can automatically intake specimen and locate target.It features fast
analyzing.It only takes 25 seconds from absorbing the specimen to displaying results.
Meanwhile, the built-in printer will print out the result,it is really quick and convenient.
n The machine implements the most advanced ACM technology in the world to
automatically monitor the operation of the solution and the status of specimen and pipes.
It can provide real time alarm as well.
n The machine can filter blister to completely remove blistering and blocking from
solution and specimens,ensuring the accuracy for the measurement.
n The equipment can provide on-line help when running ,helping users to remove
their problems.
n It is powerful in quality control function.With it ,you can have quality control
diagram and quality control statistical parameters printed.
n After conducting analysis to specimens ,the equipment will automatically flush ,
keeping the electrode standby in clean status.
n The equipment is capable of storing 1000 result,and it can be extended to be
more than 20000.The equipment will automatically refresh once it reaches the
store limit.
n The equipment features power failure protection function.It can still store
qualitycontrol and specimen data upon power failure.The storing time exceeds
5 years.
n The equipment can work 24 hours a day.If there is no operation for a long time,
the equipment will change into standby state,making it accessible at any time.It
is quite suitable for large/middle size hospitals of menu emergency calls and
specimens.
n The equipment can be powered off at any time and greatly save reagent .It
is also suitable for private hospital of county and town level since two
hospitals are of less specimen quantities.
Specifications:
n Range and accuracy of measurement:
Range of Measurement Resolution Accuracy of measurement
(CV%)
K+ 0.5~15.0mmol/L 0.01mmol/L ≤1.0%
Na+
30~200mmol/L 0.1mmol/L ≤1.0%
Cl-
20~200mmol/L 0.1mmol/L ≤1.0%
Ca++
0.1~6.0mmol/L 0.01mmol/L ≤1.5%
Li+
0.2~5.0mmol/L 0.01mmol/L ≤2.0%
pH 4~9pH 0.01pH ≤0.5%
TCO2
2.0~70.0mmol/L 0.1mmol/L ≤3.0%
n Applicable specimens:blood serum,blood p;asma,whole blood,cerebrospinal
fluid and dilute urine
n Quantity of specimens:60~300ul(item3 to item 10)
n Ambient temperature:5~35℃
n Velocity for measurement:≤25S
n Requirement on power supply:220Vą10V 50Hz60Hz
n Power rate:400W